Bruker Avance 300

Check the queue on theÌýÌýinstrument Ìý (User ID: supervisor's initials)

A broadband fluorine observe (BBFO) double resonance probeÌýdesigned for X-nuclei observation with 1H decoupling and for 1H observation is used on a permanent basis since July 2022. The X-nuclei broadband frequencyÌýrange coversÌý97Mo-31P and additionally includesÌý19F (i.e. 1H observation with 19F decoupling and vice versa areÌýpossible).

The sensitivity specifications are:Ìý

1HÌý 161 : 1 (noise [200 Hz], 0.1% Ethylbenzene in CDCl3);
13C 100Ìý: 1 (noise [40 ppm], ASTM, 40% dioxane, 60% C6D6);
15NÌý 19 : 1 (noise [2 ppm], 90% formamide in DMSO-d6);
19FÌý 344 : 1 (noise [1Ìýppm], 0.05% TFTÌýin CDCl3);
Ìý31P 120 : 1 (noiseÌý[5 ppm], 0.0485 M Triphenyl Phosphate (TPP) in acetone-d6).

The 300 MHz instrument is equipped with a BACS60 sample changer.Ìý

View of the 300 MHz instrument